The SENTEK range of standard mono ion selective electrodes are ideal for applications where high accuracy is required, particularly where ion levels are low (<1ppm)

 

 

sentekuk@btconnect.com                 +44 (0) 1376 340 456

SENTEK LIMITED

Manufacturer’s of Electrochemical Sensors

Mono ISE’s (Require Separate Reference)

Order

No.

Description

Concentration Range (Mol/L)

Lower Limits (ppm)

Temp Range oC

Ref. Elec.

Main Interference’s

pH Range

ISAB

334-75

Ammonium (NH4+)

10-1 - 10-6M

0.02

0-50

R2

K+=1.2X10-1, Na+2.0X10-3, Mg2+=2.5x10-2

5-8

4M LiAc

312-75

Barium (Ba2+)

1M - 5x10-6M

10

0-50

R2

Na+=4x10-4, K+=9x10-3, Ca2+=2.5x10-2

5-9

4M LiAc

302-75

Bromide (Br)

1M - 5x10-6M

0.4

0-80

R2

I-, S2-, CN- must be absent, OH-=3x10-5, Cl-=2.4x10-2

2-12

5M NaNO3

309-75

Cadmium (Cd2+)

10-1 - 10-6M

0.2

0-80

R1

Ag+, Hg2+, Cu2+, <10-7M

3-7

5M NaNO3

310-75

Calcium (Ca2+)

1M - 5x10-7M

0.02

0-50

R2

Mg2+, Ba2+, Pb2+, Zn2+, Na+

4-9

4M KCl

301-75

Chloride (Cl-)

1M - 5x10-5M

1.8

0-80

R2

Br, I-, CN- must be absent, S2- must be less than 10-7M

2-11

5M NaNO3

306-75

Copper (Cu2+)

1M - 5x10-6M

0.3

0-80

R2

S2-, Ag+, Hg2+ should be absent, Cl-, Br- interfere

0-7

5M NaNO3

304-75

Cyanide (CN-)

10-2 - 10-6M

0.03

0-80

R2

S2- must be <10-7M, -=1.0

10-14

5M NaOH

333-75

Fluoride (F)

1M - 5x10-7M

0.01

0-80

R1

OH- = 10-1

5-4

TISAB

303-75

Iodide (I-)

1M - 10-7M

0.02

0-80

R2

S2- =10-1 must be <10-7M/CN=1.0

3-12

5M NaNO3

307-75

Lead (Pb2+)

10-1 - 5x10-6M

1.0

0-80

R2

S2-, Ag+, Hg2+ should be absent, Cd2+, Cu2+, Fe3+ interfere

0-9

5M KNO3

311-75

Nitrate (NO3-)

1M - 5x10-6M

0.08

0-50

R2

Cl-, NO2-, Br. SO42-, F-, ClO3-, ClO4-

3-10

1M KH2PO4

314-75

Potassium (K+)

1M - 10-6M

0.04

0-50

R2

Na+, Ca2+, Rb+, Mg2+, Cs+, NH4+

4-9

TEAC

308-75

Silver (Ag+)

1M - 10-7M

0.01

0-80

R2

S2-, Hg2+, must be absent

2-9

5M NaNO3

315-75

Sodium (Na+)

Sat. - 10-6M

1ppb

-5 - +70

R2

Li+, K+, NH4+/Ag should be absent

2-12

SISAB

305-75

Sulphide (S2+)

1M - 10-7M

0.003

0-80

R1

Hg2+, Ag+ must be absent

12-14

SAOB

331-75

Voltametric Ind.

-

-

0-80

R1

-

-

-