SENTEK LIMITED

Manufacturer’s of Electrochemical Sensors

sentekuk@btconnect.com                 +44 (0) 1376 340 456

SENTEK manufactures two types of dissolved oxygen electrodes one galvanic, and the other polarographic. Both are designed for laboratory measurement and have a standard 12mm body diameter. In air-saturated water (20.9% Oxygen), the polarographic one has an output current of 600nA. The output of the galvanic electrode at saturation is 33-40mV.

The residual output in zero oxygen solutions for all the electrodes is less than 1% of the output in saturated water.

 

One or two A.T.C. sensors (Thermistor or Pt100/1000) may be fitted to customer specific requirements. Replacing the Teflon membrane is quick and easy. Simply unscrew the old assembly, load the now one with fill solution and screw on to the electrode.

The electrode is supplied complete with 2 membrane assemblies and a 50ml bottle of filling solution.

 

A 50ml bottle of Zero Oxygen solution is available for calibration purposes.

When ordering replacement electrodes please specify make and type of instrument for which the electrode is required.

Text Box: When ordering please specify:

A = Length of cable required in metres	B = A.T.C element (Th=Thermistor -
						Pt=Platinum Resistance Element)

C = Resistance value of A.T.C			D = Connector Type

e.g.	601/2m/Th10K/5Pin DIN	=	601 with 2m cable, 10K Thermistor A.T.C. and 
						5 pin DIN Connector

Type

Output at Saturation

Output at Zero Oxygen

Polarisation Voltage

Membrane

601/ABCD Polarographic

600nA ±25%

<1% Saturation

800mV

Teflon

603/ABCD Galvanic

33-40mV

<1% Saturation

N/A

Teflon